Temperature Accuracy and Precision

AIM600 Digestion Block 
 
The following data shows the accuracy and precision of the AIM600 Block Digestion System across its
operating temperature range (ambient to 4500C). The temperature specification for the AIM600  is to
maintain a consistent temperature across the block (within +/-20C). Note the performance of the block has been optimized for the situation when it is fully loaded with a full set of digestion tubes and samples.
   
It will perform slightly differently when not loaded with any digestion tubes due to the difference in thermal mass.

1. Temperature Precision at Different Temperatures

 

Temperature Precision at Different Temperatures Temperature Precision at Different Tempertures
The above graphs shows heating a fully loaded block to 800C. The block very quickly heats to 79.50C and then takes a few minutes to reach final temperature. Once the block reaches the setpoint temperature, it maintains it  within +/- 0.20C.

Heating an unloaded block to 4500C shows a similar performance though in this case there is an overshoot in the setpoint  temperature of the block caused by  reduced thermal mass of not having digestion tubes in place. As above, once the block  reaches the setpoint temperature, it  maintains the setpoint temperature within  +/- 0.20 C.

2. Precision of Temperature Ramping

 

Positive Ramp at 5 degrees per minute.

Precision of Temperature Ramping

Precision of Temperature Ramping

 

 

Positive Ramp at 1 degrees per minute.

Precision of Temperature Ramping

Precision of Temperature Ramping

The results from testing show that the control system controls the block temperature to within +/- 0.20C. This is a full order of magnitude better than the specified +/- 20C. 

Rise time is as fast as allowed by the thermal mass of the BD50s block, and is much faster than the BD50 Block Digestion System, which used a gain scheduling approach to reduce power as the block reaches a setpoint. There was no significant phase delay measured in the ramp rates.

3. Accuracy of Block Temperature

 

3.1 Calibration Accuracy

Using the thermocouple simulator the block was calibrated for 2 points (user calibration) and 5 points  (factory calibration).

The calibration values were:

 

 
Number of Points Gain Offset
2 0.6788 0.6766
5 121.56 120.715

 

 

 

After calibration, the full range of temperature from 0 to 400C was reported correctly to +/- 10C, regardless of the number of calibration points. The readings did not drift over a test period of 24 hours.

3.2 Absolute Accuracy

In section 1, it has been shown that the control accuracy is to within 0.20C. However this assumes the  thermocouple is measuring the actual block temperature. To check the thermocouple reading against the  actual block temperature, several vial reservoirs were coated in heat-sink compound, and a welded tip  thermocouple was inserted into the heat sink. The block was heated to certain temperatures according to the thermocouple and soaked for 20 minutes. The actual block temperature was measured as an average across 5  tubes spread across the block surface.

 

Controller Temperature Reading Average Block Temperature
22.0 degrees 22.7 degrees
80.0 degrees 79.2 degrees
200.0 degrees 200.6 degrees





 

These readings show the measured temperature matches the actual block temperature to within 10C.

It should be noted that these measurements are an average of 5 vial positions. All positions were within
20C of each other. Also the measurements were made near the bottom of the vials (1/3rd from the
bottom), the temperatures at the top of the vials are typically 1- 20C cooler. Therefore the specified +/- 20C has been achieved, but will depend greatly on where and how the block is measured.

4. Sample Temperature
An AIM600 Block Digestion System was fully loaded with 50 digestion tubes of water. The tubes were
surrounded with a draft shield and the block was heated to 800C. When the block reached 800C, the
water samples had only reached a temperature of 410C, indicating the delay in heat transfer from block  to sample.


 

The BD50s block was allowed to soak at 800C degrees for 1 hour. After this time, the temperature of the
samples had essentially equilibrated. The average water temperature was 71.70C.
The hottest measurement was 74.60C and the lowest was 68.50C with a std dev of 1.250C. The  distribution of temperature is shown on the surface plot below…
 

Temperature of individual Samples vs. position in the block

 

 

 

 

 

 

 

 

 

 

 
This plot shows the temperature of each sample versus its position in the block and shows there are no  localized hot or cold spots.

5. Temperature Linearity
The design of the BD50s includes a thermocouple transmitter that performs some electrical linearization on  the thermocouple signal. The firmware can then scale and shift this linear signal into an accurate temperature reading.

Seal BD 50 - Block Digestor Temperature Linearity






















The graph shows the actual block temperature versus temperature reported by the controller.

The results show the output of the transmitter is very linear across the operating range of the block and requires zero/gain compensation (linear).

6. Summary
The AIM600 controller is capable of controlling the temperature of the AIM600 block to within 0.20C.
The variation in temperature across the AIM600 block at a specific temperature is within +/- 10C.
The AIM600 block is linear in temperature across the range of ambient to 4500C.